Calibration
IMPORTANT: Only use the calibration weights provided or OIML
Class weights.
Always perform the calibration procedure AFTER the unit has been
allowed to properly warm up (at least 30 minutes). It is also recommend-
ed to recalibrate the unit periodically during long periods of use and to
always start a procedure by taring the unit to re-zero the balance.
Calibration is always recommended when first using the balance or after
the unit has been moved or subjected to an extreme change in tempera-
ture.
Calibration Procedure
1. After allowing unit to warm for 30 minutes, remove all items from the
pan and tare the balance.
2. After observing a stable “0” with an arrow in the (g) gram mode, press
and hold the CAL/MODE button until calibration weight value appears
on the display.
3. Gently place the correct calibration weight onto the pan.
4. The + sign will disappear. Wait approximately ten seconds as the unit
performs internal calibration.
IMPORTANT - Do not disturb the balance during this time. Avoid
vibration and air currents.
5. When calibration is complete the unit will beep, correct calibration
weight value will disappear for one second then reappear as an active
weight value. The balance is now calibrated.
6. Remove calibration weight and press the TARE button to reset the zero
point.
If calibration weight value remains on the LCD display, an improp-
er calibration weight has been used. Remove improper weight
and repeat calibration procedure with correct weight.