Calibration/Adjustment
The balance offers you various
calibration/adjustment functions.
You can select these functions by
setting the appropriate menu code.
All models have a built-in calibration
weight.
Using Verified Balances as Legal
Measuring Instruments in the EU*
Before using your balance as a legal
measuring instrument, you must
perform “Internal Calibration” at the
place of installation after the warmup
period.
Menu code selection: 1 9 3 **
The calibration weight is internally
applied by servomotor and removed
at the end of calibration.
Calibration is the determination
of the difference between the weight
readout and the true weight (mass) of
a sample. Calibration does not entail
making any changes within the
balance.
Adjustment is the correction of this
difference between the measured
value displayed and the true weight
(mass) of the sample, or the reduc-
tion of the difference to an allowable
level within maximum permissible
error limits.
During calibration, the sensitivity
of the balance is adjusted to changes
in the ambient conditions.
Calibrate your new balance at
the place of installation after each
warmup period and before the first
weight measurement. In verified
models the balance should be
calibrated at least once a day.
When the display shows a zero
readout, press the CAL key to activate
calibration.
If any interference affects calibration,
the error message “Err 02” is
displayed briefly.
In this case, tare and press the CAL
key again.
Important Note
During calibration, the weighing pan
must be unloaded.
** including the Signatories
of the Agreement on the European
Economic Area
** = factory setting