Problem Causes Solution
No segments No AC power Check the AC power
appear on the is available. supply.
weight display The AC adapter is not Plug in the AC adapter.
plugged in.
No segments The surface on which Make sure that ambient
appear on the weight balance rests is not stable. conditions are stable.
display after The balance has not Prevent vibrations
calibration or yet internally stabilized. affecting the surface
adjustment on which the balance
rests.
Close the draft shield.
The weight display The load capacity Unload the balance.
shows "H" exceeds the capacity
of the balance.
The weight display The weighing pan Position the pan.
shows "L" is not in place.
or "Err 54"
The weight Too much vibration Access the menu
readout changes or the balance to select the correct
constantly is exposed to a draft. code for the weighing
The draft shield is not Close the draft shield.
completely closed.
A foreign object is Remove the foreign
caught between the pan object.
and the balance housing.
The below balance Close the port for
weighing port is open. below-balance
The balance does not weighing.
have a stable weight
(absorbs moisture or
evaporates).
The sample is
electrostatically
charged.
The weight readout The balance has not Calibrate the balance.
is obviously wrong. been calibrated. tare before weighing.
The balance has not
been tared.
"Err 30" RS232 is blocked. Confirm proper pinouts
on cable.
Check selected serial
settings on balance and
attached devise.
25
Troubleshooting